Power supply |
Adjustable range |
With two phase power supply 380V 340V~420V |
With power supply 220V 195V~250V |
Fluoroscopy |
Adjustable range
of tube voltage |
50~90kV (adjustment continuously) |
Adjustable range
of tube current |
The upper limit value is not more than 5mA, the lower limit value is less than 1/10 of the upper one (adjustment continuously) |
radiography |
Adjustable range
of tube voltage |
50kV~90kV ( adjustment continuously) |
Adjustable range
of tube current |
Small focus: 15mA
Big focus: 30mA, 60mA and 100mA |
Tube current
of spot film |
100mA |
Adjustable range
of time |
0.04~6.3s 23 grades together, |
High-voltage
Generator unit |
Capacity |
9kVA instant 0.9kVA continuous |
Max DC output voltage |
90kV |
Max DC output current |
100mA instant 5mA continuous |
X-ray tube unit |
Model |
XD4-2.9/100 |
Focus |
1.8mm×1.8mm 4.3mm ×4.3mm |
|
rotary angle
|
+90°~0°~-5° |
Movable range of spot Film device |
Focus---fluorescent screen 730mm(fixed) : |
Horizontal move left(right) ≥±80mm |
Longitudinal direction: ≥500mm |
Diagnostic bed
|
Collimate for fluoroscopy |
Manual single blade |
Distance between focus and bed surface: |
At fluoroscopy and radiography of Gastrointestinal : >300mm |
Ray-filter device
|
Grid density |
N28 |
Converging distance |
f0 80 cm ﹡f0 100 cm |
Grid ratio |
r8 |
Distance between bed face and film |
≤80mm |
Move vertically along ground track |
1500mm |
Move horizontal motion of X-ray tube horizontal arm |
360mm |
Distance of move up and down |
720~1720mm (from tube focus to ground) |