Standard Composition: Servo Control Mode: air temperature, skin temperature, display humidity;
Instant Parameter Displaying: set temp, air temp , skin temp , internal humidity , heat power rate;
Audible & Visual Alarms: power failure, sensor failure, higher & lower temp, air & skin temp override, fan failure, system failure;
>37°C, Preset Mode: power failure reset memory , alarm silence, keypad lock , self-test function;
Hood with dual slop; 6 hand – ports; left & right cabinets, rail;
RS-232 Output Connector (use for network system).
Optional Devices (At Buyers’Selection):
Optional Devices (At Buyers’ Selection):
YEC—1002 Infant scale , to weigh up infant weight.
YC Stepless Adjusting Infant Bed. Adjust tilt +10°.
GL Blue Light Phototherapy Unit, LED Digital Timer, Display therapy time & total time.
SC Double Wall Canopy
|
B--3000 |
B--2000 |
B-1000 |
B-800 |
Mattress to Hood |
45mm |
450mm |
450mm |
410mm |
Observation side of hood |
Two incline |
Two incline |
Two incline |
Two incline |
Operate port |
6(2 Iris ports) |
6(2 Iris ports) |
6(2 Iris ports) |
6(2 Iris ports) |
Tubing grommets |
4 |
4 |
4 |
4 |
Front access panel |
826*310mm |
826*310mm |
826*310mm |
826*310mm |
Mattress Tray Size |
630*630mm |
630*630mm |
630*630mm |
630*630mm |
Technical Specifications |
|
|
|
|
Power requirement |
≤400VA |
≤400VA |
≤400VA |
≤400VA |
Temperature control mode |
Air and skin servo control |
Air and skin servo control |
Air servo control |
Air servo control |
Air Temp Control Range |
20℃--37℃ |
20℃--37℃ |
20℃--37℃ |
20℃--37℃ |
Temperature override mode to |
20℃--39℃ |
20℃--39℃ |
20℃--39℃ |
------ |
Sin Temp Control Range |
20℃--37℃ |
20℃--37℃ |
----- |
------- |
Temperature override mode to |
30℃--38℃ |
30℃--38℃ |
------ |
-------- |
Skin temp sensor accuracy |
≤0.3℃ |
≤0.3℃ |
------ |
------- |
Humidity display range |
5%--99% |
----- |
------- |
-------- |
Temperature fluctuation |
≤0.5℃ |
≤0.5℃ |
≤0.5℃ |
≤0.5℃ |
Mattress temperature uniformity |
≤0.8℃ |
≤0.8℃ |
≤0.8℃ |
≤0.8℃ |
Temperature rise time |
≤30 min |
≤30 min |
≤30 min |
≤30 min |
Noise level within hood environment |
≤55dB |
≤55dB |
≤55dB |
≤55dB |
Alarms |
|
|
|
|
Power failure alarm |
Yes |
Yes |
Yes |
Yes |
Fan failure alarm |
Yes |
Yes |
Yes |
Yes |
Sensor failure alarm |
Yes |
Yes |
Yes |
Yes |
High air temperature alarm |
+1.5℃ |
+1.5℃ |
+1.5℃ |
+1.5℃ |
Low air temperature alarm |
-2.5℃ |
-2.5℃ |
-2.5℃ |
-2.5℃ |
High skin temperature alarm |
+1.0℃ |
+1.0℃ |
----- |
------ |
Low skin temperature alarm |
-1.0℃ |
-1.0℃ |
----- |
----- |
Over air temperature alarm |
≤38.0℃;
≤40.0℃ |
≤38.0℃;
≤40.0℃ |
≤38.0℃;
≤40.0℃ |
≤38.0℃; |
Over skin temperature alarm |
≤38.0℃;
≤39.0℃ |
≤38.0℃;
≤39.0℃ |
≤38.0℃;
≤39.0℃ |
----- |
System failure alarm |
Yes |
Yes |
Yes |
Yes |
Alarm silence function |
Yes |
Yes |
Yes |
Yes |
Operational condition |
|
|
|
|
Environment temperature |
+20℃-+30℃ |
+20℃-+30℃ |
+20℃-+30℃ |
+20℃-+30℃ |
Environment relative humidity |
5%---99% |
5%---99% |
5%---99% |
5%---99% |
Environment air velocity of flow |
<0.3m/s |
<0.3m/s |
<0.3m/s |
<0.3m/s |
others |
|
|
|
|
RS—232 Connector |
Yes |
Yes |
Yes |
Yes |
Rail |
Yes |
Yes |
Yes |
Yes |
Infant bed tilt adjusting |
±10°step-less |
±10°step-less |
±10°step-less |
----- |
Mattress tray Max. load |
10kg |
10kg |
10kg |
10kg |
Transfusion pole Max. load |
2kg |
2kg |
2kg |
2kg |
Cabinet Max. load |
6kg |
10kg |
10kg |
----- |
Packing size (2cartons) |
100*58*80cm |
100*58*80cm |
100*58*80cm |
100*58*80cm |
Gross Weight |
120kg |
120kg |
120kg |
120kg |
|